• Ql-300b 99.9999% Purity Lab Use Hydrogen Generator for Gas Chromatography
  • Ql-300b 99.9999% Purity Lab Use Hydrogen Generator for Gas Chromatography
  • Ql-300b 99.9999% Purity Lab Use Hydrogen Generator for Gas Chromatography
  • Ql-300b 99.9999% Purity Lab Use Hydrogen Generator for Gas Chromatography
  • Ql-300b 99.9999% Purity Lab Use Hydrogen Generator for Gas Chromatography
  • Ql-300b 99.9999% Purity Lab Use Hydrogen Generator for Gas Chromatography

Ql-300b 99.9999% Purity Lab Use Hydrogen Generator for Gas Chromatography

Appliance: Gc
Ray Type: Gc
Gaspath System: Single
Detector: FID
Energy Resolution: ≤7%
Certification: CE, ISO
Samples:
US$ 890/Piece 1 Piece(Min.Order)
| Request Sample
Customization:
Manufacturer/Factory & Trading Company
Gold Member Since 2019

Suppliers with verified business licenses

Shandong, China
to see all verified strength labels (27)

Basic Info.

Model NO.
QL-300B
Chromatography Type
Gas Chromatography
Chromatographic Column Type
Convetional Analytical Column
Chromatographic Column Length
10~30cm
Chromatographic Column Diameter
0.2~0.5mm
H2 Flow Rate
300ml/Min
H2 Purity
99.999%
H2 Pressure
0-4 Bar or 0-7 Bar
Transport Package
Wooden Box
Specification
H2 flow: 300ml/min, purity: 99.999%
Trademark
Saikesaisi
Origin
China
HS Code
8543300090
Production Capacity
300unit/Month

Product Description

QL-300 ultra-pure hydrogen generator   

QL-300 hydrogen generator adopt SPE/PEM technology-top hydrogen tech in the world. PEM full name is Proton Exchange Membrane. Also called SPE technology ( solid polymer electrolyte).QL series hydrogen generator is our patented products, which are light , highly effective  energy-saving and of environmental protection, producing extremely pure Hydrogen (>99.9999%) through the electrolysis of pure water.

 

Application:

1.GC (gas chromatography) fuel gas and carrier gas(replace h2 cylinder)

2.ELCD (conductivity detector) reaction h2 gas(replace h2 cylinder)

3.AED (atomic emission detector) of the reaction h2 gas(replace h2 cylinder)

4.Pure h2 gas supply for fuel cell

5.Other pure h2 using field

 

Technical Parameter:

Type QL-300
Output Volume (ml/min) 0~310

Output Pressure (MPa)

0.02~0.4

Pluctuation Rate of Output Pressure(%)

<0.1

H2 purity (%)

>99.999

Second Protective Pressure (MPa)

0.42

Power Voltage (V)

220±15%   110±15%    50~60Hz

Input power (W)

<150

Dimension (LxWxH) mm

431x228x343

Gross Weight (KG)

< 15

 

PEM hydrogen Technology working principle:

The proton exchange membrane (PEM), which only allows water and positive ions to cross between compartments. The membrane also serves as the electrolyte in the cell, eliminating the need for hazardous liquid electrolytes such as concentrated potassium hydroxide. PEM water electrolysis simply splits pure deionized water (H2O) into its constituent parts, hydrogen (H2) and oxygen (O2), on either side of this membrane. When a DC voltage is applied to the electrolyzer, water fed to the anode, or oxygen electrode, are oxidized to oxygen and protons, while electrons are released. The protons (H+ ions) pass through the PEM to the cathode, or hydrogen electrode, where they meet electrons from the other side of the circuit, and are reduced to hydrogen gas. The two reactions that occur in the cell are as follows:
2.2H2O -> 4H+ + 4e- +O2
3.4H+ + 4e- -> 2H2
4.Thus, the only possible components of the streams are hydrogen, oxygen and water moisture, as shown in 
Figure 1 below.

Ql-300b 99.9999% Purity Lab Use Hydrogen Generator for Gas Chromatography
 

 

Characteristics of the Products:

1.Electrolysis of pure water (adding no alkali) to produce hydrogen with no corrosion and pollution, and the purity of hydrogen is extremely high.

2.Voltage of unit cell is low, and electric resistance in the electrolysis cell is little, giving out no heat, and cycle of changing desiccants is long, producing highly pure hydrogen.

3.With stable pressure and output, Hydrogen is controlled according to the requirements, and automatic control technology is perfect and reliable.

4.Power consumption is less, and the electrolysis is efficient.

5.H2 flow and pressure is stable

6.Titanium material

7.DuPont Nafion Membranes N-117

Ql-300b 99.9999% Purity Lab Use Hydrogen Generator for Gas Chromatography
Ql-300b 99.9999% Purity Lab Use Hydrogen Generator for Gas Chromatography
Ql-300b 99.9999% Purity Lab Use Hydrogen Generator for Gas Chromatography
 

 

Packaging & Shipping

Ql-300b 99.9999% Purity Lab Use Hydrogen Generator for Gas Chromatography
 

Ql-300b 99.9999% Purity Lab Use Hydrogen Generator for Gas Chromatography
 

 

Rick Hu
Shandong Saikesaisi Hydrogen Energy Co.,Ltd
 

 
 
 
NO.1999 Dazheng Road, Gaoxin East District,Jinan,China 250104

Send your message to this supplier

*From:
*To:
*Message:

Enter between 20 to 4,000 characters.

This is not what you are looking for? Post a Sourcing Request Now